The shape and fitting position of a silicone rubber gasket vary depending on the type of an injection holder. When replacing the silicone rubber gasket, refer to the following instructions and use the procedure suitable for the model of your injection holder.
* The HI-7 comes standard with the IM-9A/9B/9C as well as the IM-6/IM-5A/IM-5B/IM-6-2 sold after 2000.
HI-7 | |
---|---|
Remove the injection holder cap and replace the worn-out silicone rubber gasket in the injection holder with a new one, and then put the injection holder cap HIC back on. (If HIC is lost or missing, obtain a new injection holder cap HIC.) |
HI-6 (Ver.1) | |
---|---|
Remove both the end cap and the stainless steel collar from the injection holder and replace the worn-out silicone rubber gasket in the end cap with a new one, and then put the stainless steel collar and end cap back on. (Please note that both the end cap and the stainless steel collar used for HI-6 Ver.1 are no longer available as replacement parts. If they are lost, obtain a new injection holder cap HIC.) |
HI-6 (Ver.2) | |
---|---|
Remove both the end cap and the stainless steel collar from the injection holder and replace the worn-out silicone rubber gasket in the end cap with a new one, and then put the stainless steel collar and end cap back on. (Please note that both the end cap and the stainless steel collar used for HI-6 Ver.2 are no longer available as replacement parts. If they are lost, obtain a new injection holder cap HIC.) |
HI-6 (Ver.3) | |
---|---|
Remove both the end cap and the stainless steel collar from the injection holder and replace the worn-out silicone rubber gasket in the injection holder with a new one, and then put the stainless steel collar and end cap back on. (Please note that both the end cap and the stainless steel collar used for HI-6 Ver.3 are no longer available as replacement parts. If they are lost, obtain a new injection holder cap HIC.) |
HI-4 (Ver.1) | |
---|---|
Remove the end cap from the injection holder and replace the worn-out silicone rubber gasket with a new one, and then put the end cap back on. (Please note that the end cap used for HI-4 Ver.1 is no longer available as replacement parts.) |
HI-4 (Ver.2) | |
---|---|
Remove both the end cap and the stainless steel collar from the injection holder and replace the worn-out silicone rubber gasket in the injection holder with a new one, and then put the stainless steel collar and end cap back on. (Please note that both the end cap and the stainless steel collar used for HI-4 Ver.2 are no longer available as replacement parts. If they are lost, obtain a new injection holder cap HIC.) |